9-2. F-Functions

‰ “ ” designates Factory Settings.

F-00-X Weight Unit

USA Version ONLY

f-00-01

‰Weight Display when the scale is switched on. See also setting “f-09-01=2”.

0

kg (kilograms).

 

 

1

lb (pounds).

 

 

 

‰ Unit Weight (when “lb” is selected).

 

 

 

 

f-00-02

 

0

lb as piece weight.

 

 

 

1

lb as 1,000 piece weight.

 

 

 

F-01-X Operations

f-01-01

‰Operation Mode.

0

Normal Operation. All features and keys available.

 

 

1

Simplified Operation. The unit weight registration is by

sample only. All other keys are disabled.

 

f-01-02

‰“Add” Sample Request Override.

If the sample weight is too light and the scale asks to “Add” more sample pieces, using this F-Function, the unit weight can be entered without adding the requested sample pieces. Or disable the “Add” Sample Request function.

0

“Add” sample request function is disabled. Light unit weight

can be accepted without “Add” more sample request.

 

1

The unit weight can be entered without requested “Add”

sample pieces (via the

ENTER

key).

 

2

The unit weight cannot

be entered without requested “Add”

sample pieces (via the

ENTER

key).

 

‰“f-01-03” is for factory use and should be “0”.

f-01-04

‰Display ON Unit Weight – Reset or Last.

 

When the scale is switched on, the scale can be set to recall

 

the last unit weight used.

0

 

The unit weight is RESET (cleared) when display comes on.

 

 

 

1

 

The unit weight last used will be entered automatically.

31

Page 33
Image 33