Baud Rate (f-06-04)

Select the baud rate according to the device to be connected.

‰2400 bps (f-06-04="0") Select 2400 bps to connect with an AD-8121.

‰4800 bps (f-06-04="1")

‰9600 bps (f-06-04="2")

13-4. Connecting the AD-8121 Printer / MODE 1 or MODE 2

‰When using the AD-8121 printer (MODE 1 or MODE 2), you will be able to get data: Number of data items, total, maximum, minimum, mean value, range of data (max. - min. data) and standard deviation.

‰When using the AD-8121 with MODE 2, set f-06-02 to print pcs (count) data only or weight data only.

‰To print date and time, use the AD-8121’s calendar / clock function and set f-06-02 to print pcs (count) data only or weight data only.

Print Operations Settings

Print By:

 

F-Function f-06-01

Printer MODE

HC

 

 

 

key

 

MODE 1

PRINT

0

 

 

 

 

 

Auto Print

 

2 or 3

MODE 1

Printer

 

 

key

 

MODE 2

DATA

1

 

 

 

 

 

 

 

Example of f-06-02settings

‰To print pcs (count) data only: set f-06-02at "0100"

‰To print weight data only: set f-06-02at "0010"

‰To print pcs (count) and weight data: set f-06-02at "0110"

‰To print pcs, weight and unit weight data: set f-06-02at "0111"

‰To print total data (accumulated by the M+ key), press the TOTAL count display shows the total, then press the PRINT key.

key so the

‰If you are using the AD-8121’s statistic functions, then set f-06-02at "01#0" (# = 0 or 1) for pcs (count) data or "0010" for weight data.

‰MODE 1 and 2 of the AD-8121 can not print ID numbers.

45

Page 47
Image 47
A&D HC-6Ki, HC-15Ki, HC-3Ki Connecting the AD-8121 Printer / Mode 1 or Mode, Baud Rate f-06-04, Print Operations Settings