36
Data Length and Parity
07 bits, even parity.
17 bits, odd parity.
28 bits, non parity.
f-07” and “f-08” are for factory use and should be “0”.

F-09-X ½ key

Operation mode for the ½key
0Operating as a key to display and/or change comparator
upper and lower limits.
1Operating as M- key to subtract Count data from M+
memory.
2Operating as a key to toggle the weighing unit between “lb”
and “g”.
Decimal Point and RS-232C Output
0000Decimal point: “.
Header for stable weight data: “ST”
Acknowledgment of command: <ACK><CR><LF>
Select the decimal point “.” or ” ,” and output
format for RS-232C. Set 0 or 1 for each bit.
Decimal point: “0” = ”.
“1” = ”,
Header: “0” = “ST,+001.2346 kg”
“1” = “WT,+001.2346 kg”
Acknowledgment: “0” = ”<ACK><CR><LF>”
“1” = ”<ACK>”
f-10”, “f-11” and “f-12” are for factory use and should be “0”.
f-06-05f-09-01
F
-
0
9
-
01=2 is for
USA version
only.

0000

ST/WT
N.A.
Decimal
ACK Point
f-09-02