13. OP-03 RS-232C SERIAL INTERFACE

This interface allows the HC-iseries to be connected with a multifunction printer or a personal computer.

‰The OP-03 unit includes an interface board, a connector plug (DIN type) and two screws. (M3x6 tapping type).

13-1. Installation

1.Disconnect the AC adapter from the scale. If the battery is used, switch off the scale.

2.Loosen the screw and remove the panel covering the option slot.

3.Connect the connector cable on the OP-03 to the connector inside the option slot.

4.Fix the OP-03 unit using the two screws included in the OP-03.

Option slot

Interface

 

Screw

 

Screws

13-2. RS-232C Specifications

 

Transmission form

Asynchronous, bi-directional, half-duplex

Data format

Baud rate:

2400, 4800, 9600 bps

 

 

Data:

7 bits + parity 1bit (even / odd) or 8 bits (non-parity)

 

Start bit:

1 bit

 

 

Stop bit:

1 bit

 

 

Code:

ASCII

 

 

Terminator:

Data Send / CRLF

Data Receive / CR or CRLF

LSB

0 1 2 3 4 5

MSB

6

1(-15V~-5V)

0(5V~15V)

Start bit

 

Data bit

 

Parity bit

 

 

Stop bit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pin connections

(View from outside)

Mating connector: JA+TCP0586 (Included in the OP-03)

2Receive data

3Transmit data

5Signal ground

7Data set ready

1, 4, 6 and 8 N.C.

Inside of the HC-i

43

Page 45
Image 45
A&D HC-3Ki, HC-15Ki, HC-30Ki, HC-6Ki manual OP-03 RS-232C Serial Interface, Installation, RS-232C Specifications, Ascii