Agilent Technologies 82357A manual Software Installation Checks, Check Suspend/Resume Operation

Models: 82357A

1 70
Download 70 pages 54.11 Kb
Page 49
Image 49

Troubleshooting the 82357A

Software Installation Checks

Software Installation Checks

When only the red FAIL LED is still ON after 10 seconds, the 82357A has been detected by the host computer, but has not yet been configured for use with the Agilent IO Libraries. Possible causes for this is that the appropriate version of the Agilent IO Libraries or Agilent IO Libraries Suite has not been installed on your PC or the 82357A USB drivers have not been installed.

Check Suspend/Resume Operation

Some Windows operating systems support Power Management which can suspend the PC while the 82357A is in operation. After a Suspend/Resume cycle, the 82357A may not properly resume operation. In this case, you may need to unplug/replug the USB cable to restore 82357A operation. If this does not correct the problem, go to “Verify Agilent IO Libraries Suite Installation”.

NOTE

If your 82357A applications must not be preempted by a PC Suspend event, we recommend you disable Power Management on your PC by using the Control Panel Power Options dialog.

Verify Agilent IO Libraries Suite Installation

When only the red LED is ON after 10 seconds and all Windows Plug and Play Manager activity has ceased, start your troubleshooting sequence by verifying IO Libraries installation. If any action taken results in a change in the LED status, go to “Software Configuration Checks” or to “Hardware Checks”.

1Check Agilent IO Libraries Version. If a version of the Agilent IO Libraries or IO Libraries Suite has been installed, a blue IO icon is normally displayed on the Windows taskbar (on the lower right-hand side of the screen).

IO Libraries Icon

IO Libraries Suite Icon

Chapter 3

49

Page 49
Image 49
Agilent Technologies 82357A manual Software Installation Checks, Check Suspend/Resume Operation