Setting up the HP/Agilent 8922 GSM HSCSD Test System

Making the front panel connections

Making the front panel connections

Before you connect the two HP/Agilent 8922 P systems together to form the HP/Agilent 8922 GSM HSCSD Test System, ensure that both HP/ Agilent 8922 P systems are connected and calibrated as described in the HP/Agilent 8922 Multi-Band Test System Supplementary User’s Guide. For details on the HP/Agilent 8922P front panel connections, see the HP 8922 Multi-Band Test System Supplementary User’s Guide.

The RF flow must be combined and split between the two HP/Agilent 8922P systems. This is done by connecting the supplied RF link cable (08922-61185) between the Coupled RF Port on the dual band HP/ Agilent 83220 of the master unit and the GSM/DCS/PCS RF In/Out port on the HP/Agilent 83220 of the slave unit.

Figure 2-1

Front panel connections

Master

Slave

Connect mobile here

RF link cable

8

Chapter 2

Page 22
Image 22