Remote operation

 

HSCSD Subsystem GPIB Commands

 

HSCSd:SLAVe:DOWNlink:AMPLitude:OFFSet

Description

This sets the amplitude offset (in dBs) of the slave timeslot, using the

 

master’s downlink amplitude as a reference.

 

The resulting amplitude must lie within the normal downlink power

 

range. If not, the error Disallowed due to pulse mod configuration

 

is produced, and the setting is ignored.

Syntax

HSCSd:SLAVe:DOWNlink:AMPLitude?

 

HSCSd:SLAVe:DOWNlink:AMPLitude:OFFSet <real>

Command options

Type

floating point

 

 

Units:

dB only

 

 

Limits

-100.0 thru 100.0

 

 

Default

0.0

 

 

58

Chapter 5

Page 72
Image 72