On both units, if the instrument is operating in standard mode, setting this field changes the slave control field (see page 37) to MS Tx Lev, then calculates and sets the slave expected input level (see page 36). If the instrument is operating in compatibility mode, setting this field only changes the slave expected input level when the slave control type is MS Tx Lev.

NOTEThe default is reset when you select the instrument type (see page 24).

13. Slave input level

Type

float

 

 

Units

dBm only

 

 

Limits (GSM900)

-27 thru +41.0

 

 

Limits (E-GSM)

-27 thru +41.0

 

 

Limits (DCS1800)

-47 thru +33.0

 

 

Limits (PCS1900)

-47 thru +33.0

 

 

Default (compatibility mode)

33.0

 

 

Default (GSM900)

13.0

 

 

Default (E-GSM)

13.0

 

 

Default (DCS1800)

10.0

 

 

Default (PCS1900)

10.0

 

 

The Slave input level field sets the expected power of the slave uplink TCH. You can set the level indirectly by changing the Slave Tx level (see page 35) or change it directly. This value is used to set the RF analyzer level when the measurement timeslot (see page 32) is set to Slave.

If the instrument is operating in standard mode, setting this field changes the slave control type (see page 37) to Manual.

If the instrument is operating in compatibility mode, and the slave

36