Screens

HSCSD screen

9. Master Tx level

Type

integer

 

 

Units

none (Tx level)

 

 

Limits (GSM900)

1 thru 19

 

 

Limits (E-GSM)

1 thru 19

 

 

Limits (DCS1800)

0 thru 15

 

 

Limits (PCS1900)

0 thru 15; 30 thru 31

 

 

Default (GSM900)

15

 

 

Default (E-GSM)

15

 

 

Default (DCS1800)

10

 

 

Default (PCS1900)

10

 

 

On a master unit, the Master Tx level field requests a new transmit level for the uplink master TCH. On a slave unit, this field does not make this request.

If the instrument is operating in standard mode, setting this field changes the master mode (see page 35) to MS TX Lev, then calculates and sets the master’s expected input level (see page 34). If the instrument is operating in compatibility mode, setting this field only changes the master’s expected input level when the control type for the master is set to MS TX Lev.

NOTE

The default is reset when you select the instrument type (see page 24).

 

 