Setting up the HP/Agilent 8922 GSM HSCSD Test System
Making the rear panel connections
Making the rear panel connections
Before connecting the two HP/Agilent 8922 P systems to form the HP/ Agilent 8922 GSM HSCSD Test System, ensure that both HP/Agilent 8922 P systems are connected and calibrated as described in the HP/ Agilent 8922 
To make the rear panel connections of the HP/Agilent 8922 GSM HSCSD Test System:
| Step 1. | Connect the system bus cable from the master unit to the slave unit (the | 
| 
 | end with the BNC connection must be connected to the master unit). | 
| 
 | If you are not connecting the BNC connector on the system bus cable to | 
| 
 | for example, a spectrum analyzer, then terminate the connection using | 
| 
 | the BNC terminator provided. | 
| Step 2. | Connect the GPIB cable from the GPIB port on the master unit to the | 
| 
 | GPIB port on the slave unit. | 
| 
 | If you are operating the HP/Agilent 8922 GSM HSCSD Test System | 
| NOTE | |
| 
 | manually do not connect another GPIB controller to the GPIB bus. The | 
| 
 | master unit operates as the controller and another device attached to the | 
| 
 | GPIB bus results in conflicts on the GPIB bus. | 
| 
 | 
 | 
| Step 3. | Connect the 10 MHz reference cable from the 10MHz OUT port on the | 
| 
 | master unit to the Ref IN port on the HP/Agilent 83320 of the slave unit. | 
| 
 | The HSCSD connections are shown below. | 
| Chapter 2 | 9 | 