Manual operation

Configuring the system

 

Configuring the system

 

To configure the HP/Agilent 8922 GSM HSCSD Test System for manual

 

operation, proceed as follows.

 

You must configure the slave unit first as a frame reset is sent from the

NOTE

 

master unit to the slave unit as soon as the master unit is activated. The

 

slave unit must be ready to receive this frame reset.

 

 

NOTE

If you are operating the HP/Agilent 8922 GSM HSCSD Test System

 

manually do not connect another GPIB controller such as a PC to the

 

GPIB bus. The master unit operates as the controller and any other

 

device attached to the GPIB bus can result in conflicts on the GPIB bus.

 

 

Step 1. On the slave unit, navigate to the CONFIG screen and set the Instrument Type to HSCSD Slave.

Figure 3-1

Config screen

 

 

 

 

 

 

 

 

 

 

 

 

 

C.01.06

16

Chapter 3

Page 30
Image 30
Agilent Technologies 8922 P manual Configuring the system, Config screen