Agilent Technologies 8922 P manual HSCSdMASTerDOWNlinkAMPLitude, Command 0ptions

Models: 8922 P

1 94
Download 94 pages 14 Kb
Page 64
Image 64

 

Remote operation

 

HSCSD Subsystem GPIB Commands

 

HSCSd:MASTer:DOWNlink:AMPLitude

Description

This sets the downlink amplitude of the master TCH. In all cases, this

 

sets the TCH level. To change the amplitude of the GSM BCH in a dual

 

band system, use the existing command DUALband:ATTenuation:GBCh.

 

See the HP/Agilent 8922 Multi-Band Test System Supplementary User’s

 

Guide for more details.

 

This command must also be sent to the slave unit.

Syntax

HSCSd:MASTer:DOWNlink:AMPLitude?

 

HSCSd:MASTer:DOWNlink:AMPLitude <real>

Command 0ptions

Type

floating point

 

 

Units:

dBm only

 

 

Limits (GSM900):

-127.0 thru +1.0

 

 

Limits (E-GSM):

-127.0 thru +1.0

 

 

Limits (DCS1800):

-127.0 thru -12.0

 

 

Limits (PCS1900):

-127.0 thru -12.0

 

 

Default:

-85.0

 

 

50

Chapter 5

Page 64
Image 64