NOTE

Remote operation

Remote operating procedures

Remote operating procedures

For remote operation to work:

the two HP 8922P units must be put into master and slave mode by remote control

HSCSD settings must be made on both HP/Agilent 8922 P units

Configuring the system

The slave unit must be configured and initialized before the master unit. Note that unlike manual configuration, the GPIB address of the slave unit does not have to be synchronized on both units.

!Configure instruments; note we do slave first

!for a single box, we send the single command

!selecting ‘HSCSD SINGLE’

OUTPUT Slave;”CONFIGURE:INSTRUMENT:TYPE ‘HSCSD SLAVE’”

WAIT 3 !to make sure we are ready to receive reset from master OUTPUT Master;”CONFIGURE:INSTRUMENT:TYPE ‘HSCSD MASTER’”

The slave unit must wait for the frame reset from the master unit.

! select HSCSD screen

OUTPUT Master;”DISPLAY:SCREEN HSCSD” OUTPUT Slave;”DISPLAY:SCREEN HSCSD”

!and wait for slave to receive reset from master

REPEAT !

OUTPUT Slave;”HSCSD:SLAVE:STATUS:SYNC?” ENTER Slave;sync$

UNTIL sync$=”””ENABLED”””

62

Chapter 5

Page 76
Image 76
Agilent Technologies 8922 P manual Remote operating procedures, Configuring the system