Remote operation

 

 

 

HSCSD Subsystem GPIB Commands

 

HSCSd:SLAVe:UPLink:AMPLitude

Description

This sets the expected input level of the slave timeslot.

 

In standard mode, this sets the slave’s uplink amplitude control to

 

MANUAL. In compatibility mode, if the slave’s uplink amplitude control is

 

set to MS TX LEV then an error occurs and the setting is not performed. If

 

the measurement timeslot (see page 43) is set to Slave, the analyser

 

level of the unit is set to this value.

Syntax

HSCSd:SLAVe:UPLink:AMPLitude?

 

HSCSd:SLAVe:UPLink:AMPLitude <real> <units>

Command options

 

 

 

 

 

 

 

 

Type

float

 

 

 

 

 

 

Units:

dBm only

 

 

 

 

 

 

Limits (GSM900):

-27.0 thru +41.0

 

 

 

 

 

 

Limits (E-GSM):

-27.0 thru +41.0

 

 

 

 

 

 

Limits (DCS1800):

-47.0 thru +33.0

 

 

 

 

 

 

Limits (PCS1900):

-47.0 thru +33.0

 

 

 

 

 

 

Default (compatibility mode):

33.0

 

 

 

 

 

 

Default (GSM900)

13.0

 

 

 

 

 

 

Default (E-GSM):

13.0

 

 

 

 

 

 

Default (DCS1800):

10.0

 

 

 

 

 

 

Default (PCS1900):

10.0

 

 

 

 

 