60 Chapter5
Remote operation
HSCSD Subsystem GPIB Commands
HSCSd:SLAVe:UPLink:AMPLitude
Description This sets the expected input level of the slave timeslot.
In standard mode, this sets the slave’s uplink amplitude control to
MANUAL. In compatibility mode, if the slave’s uplink amplitude control is
setto MS TX LEV thenan error occurs and the setting is not performed. If
the measurement timeslot (see page 43) is set to Slave, the analyser
level of the unit is set to this value.
Syntax HSCSd:SLAVe:UPLink:AMPLitude?
HSCSd:SLAVe:UPLink:AMPLitude <real> <units>
Command options
Type float
Units: dBm only
Limits (GSM900): -27.0 thru +41.0
Limits (E-GSM): -27.0 thru +41.0
Limits (DCS1800): -47.0 thru +33.0
Limits (PCS1900): -47.0 thru +33.0
Default (compatibility mode): 33.0
Default (GSM900) 13.0
Default (E-GSM): 13.0
Default (DCS1800): 10.0
Default (PCS1900): 10.0