Agilent Technologies 8922 P manual HSCSdMASTerTCHTLEVel

Models: 8922 P

1 94
Download 94 pages 14 Kb
Page 65
Image 65

 

Remote operation

 

HSCSD Subsystem GPIB Commands

 

HSCSd:MASTer:TCH:TLEVel

Description

During a call, this sets the Tx level of the master TCH.

 

When the system is idle (that is, not on a call), this sets the Tx level to be

 

used when the next call is established. Note that this value is used

 

initially for both timeslots. It is only once the TCH is established, and the

 

mobile begins to decode the downlink SACCH, that the individual Tx

 

levels are used.

 

In standard mode, this also sets the master unit’s uplink amplitude

 

control to MS TX LEV, and calculates and sets the master unit’s expected

 

uplink amplitude. See [reference] below for details of these commands.

Syntax

HSCSd:MASTer:TCH:TLEVel?

 

HSCSd:MASTer:TCH:TLEVel <integer>

Command options

Type

integer

 

 

Units:

none (Tx Level)

 

 

Limits (GSM900):

1 thru 19

 

 

Limits (E-GSM):

1 thru 19

 

 

Limits (DCS1800):

0 thru 15

 

 

Limits (PCS1900):

0 thru 15; 30 thru 31

 

 

Default (GSM900):

15

 

 

Default (E-GSM):

15

 

 

Default (DCS1800):

10

 

 

Default (PCS1900):

10

 

 

Chapter 5

51

Page 65
Image 65
Agilent Technologies 8922 P manual HSCSdMASTerTCHTLEVel