NOTE

NOTE

Screens

HSCSD screen

4. Master timeslot

Type

integer

 

 

Units

none (timeslot)

 

 

Limits

2 thru 5

 

 

Default

4

 

 

The Master timeslot field indicates which timeslot is used for the master TCH. As the slave TCH is always in the timeslot immediately after the master, changing this field sets the slave timeslot to the value “master timeslot plus one”.

On a master unit or a single unit, the value of this field is always the same as the TCH timeslot field on the Cell Status screen.

The default is reset when the instrument type (see page 24) is selected.

5. Slave timeslot

Type

read-only

 

 

Values

3 thru 6

 

 

Default

5

 

 

The relative position of the slave timeslot is fixed at +1. Therefore, if the master timeslot is set to 3, the slave timeslot is set to 4. Likewise, as the master timeslot changes, this field also changes.

On a slave unit, the value of this field is always the same as the TCH timeslot field on the Cell Status screen.

The default is reset when the instrument type (see page 24) is selected.

30

Chapter 4

Page 44
Image 44
Agilent Technologies 8922 P manual Master timeslot, Slave timeslot