Agilent Technologies Agilent 86120C manual If the MEASSCALPOWWNUM? 6451 command is sent, and a

Models: Agilent 86120C

1 294
Download 294 pages 52.87 Kb
Page 152
Image 152
If the :MEAS:SCAL:POW:WNUM? 6451 command is sent, and a 1550

Programming Commands

Measurement Instructions

<resolution>

MAXimum

0.01 resolution (fast update)

Constants

MINimum

0.001 resolution (normal)

 

 

DEFault

Current resolution

Examples

:CONF:ARR:POW:WNUM DEF, MAX

 

:FETC:ARR:POW:WNUM? DEF, MIN

 

:READ:ARR:POW:WNUM?

 

:MEAS:ARR:POW:WNUM?

 

:CONF:SCAL:POW:WNUM 6451, MAX

 

:FETC:SCAL:POW:WNUM? 6451, MIN

 

:READ:SCAL:POW:WNUM? 6451

 

:MEAS:SCAL:POW:WNUM? 6451

Query Response

If the :MEAS:SCAL:POW:WNUM? 6451 command is sent, and a 1550

 

nm laser line is present, the following response would be returned to

 

the computer:

 

 

+6.45286262E+005

 

Notice that the returned units are m–1. If six laser lines are measured and :MEAS:ARR:POW:WNUM? is sent, the following response is returned. Notice that the first returned number indicates the number of laser- line values returned in the query.

6,+6.47298400E+005,+6.46627900E+005,+6.45957000E+005,+6.45286300E+005,+6.44615600E+0

05,+6.43945300E+005

4-24

Page 152
Image 152