MEASure{:ARRay  [:SCALar]} :POWer:WNUMber?

Programming Commands

Measurement Instructions

MEASure{:ARRay [:SCALar]} :POWer:WNUMber?

Returns a wave number value.

Syntax

:POWer:WNUMber? [<expected_value>[,<resolution>]]

 

 

 

 

 

 

Used With

<expected_value>

<resolution>

 

 

 

 

 

SCALar

optional

optional

 

ARRay

ignoreda

optional

a. Although ignored, this argument must be present if the resolution argument is specified.

Description When used with a :SCALar command, a single value is returned. The display is placed in the single- wavelength mode, and the marker is placed on the signal having a wave number that is closest to the <expected_value> parameter. Default units for <expected_value> parameter are in m–1.

When used with an :ARRay command, an array of wave number is returned. The display is placed in the list- by- wavelength mode.

The <resolution> parameter sets the resolution of the measurement. It is a unitless number whose value will be limited to either 0.01 or 0.001 (whichever is closer).

Returned values are in inverse meters. Displayed units are inverse centimeters. Power units are not affected.

CONFigure command

When this function is used with the CONFigure command, the query question mark character “?” must not be included in the string. However, the FETCh, READ, and MEASure command are queries and require the question mark. Refer to the examples for this command.

<expected_value> MAXimum

The laser line having the largest wave number

Constants

MINimum

The laser line having the smallest wave number

 

 

DEFault

The current marker position

4-23