To measure signal- to- noise

Making Measurements

Measuring Signal-to-Noise Ratios

To measure signal- to- noise

1Press the front- panel Preset key.

2Press List by WL or List by Power.

3Press Appl’s and then S/N.

4To select the wavelength reference for measuring the noise, do the following steps:

a Press WL REF, and

press AUTO to let the instrument interpolate the wavelength, or

press USER to select the last wavelength manually entered.

bIf you chose USER, you can specify the wavelength by pressing USER

WL. Use the Manual background softkey to select the digit that requires editing. Use

the Manual background and Manual background softkeys to change the value.

cPress RETURN.

5While the signal- to- noise measurements are displayed, you can press PEAK anytime to select the signal with the highest power.