Troubleshooting the Analyzer

 

To troubleshoot fast bus failures

To troubleshoot fast bus failures

Use this test to isolate Fast Bus failures to the A7 CPU assembly or A6 Digital assembly.

Set the power switch to off ( O ).

Set the power switch to on ( l ) while holding in the [ System Utility ] key.

The screen displays Fast Bus Diagnostic Test ... and the power-on LEDs are flashing.

If the analyzer did not respond correctly, the A7 CPU assembly is probably faulty.

Using a logic probe, check the following signals.

A7 P10 Pin

Signal Name

TTL Logic State In Test Mode

 

 

 

 

 

 

64, 114, 65, 115, 66

FA1 to FA5

Toggling

 

 

 

72

ECLK

Toggling

 

 

 

74

FSELAn

Toggling

 

 

 

112

BRESETn

Low

 

 

 

119

FRW

Toggling

 

 

 

123

FIFOENn

High

 

 

 

124

FSELSn

Toggling

 

 

 

7-11, 21-45, 59, 71, 109,

GND

Low

118, 120, 122, 147, 149

 

 

If the signals are correct, the A6 Digital assembly is probably faulty.

If any signal is incorrect, the A7 CPU assembly is probably faulty.

This is only a partial check of the fast bus signals between the A7 CPU assembly and the A6 Digital assembly.

4-29

Page 205
Image 205
Agilent Technologies 35670-90066 To troubleshoot fast bus failures, A7 P10 Pin Signal Name TTL Logic State In Test Mode