To set up the source amplitude accuracy test

Agilent 35670A

Verifying Specifications

 

To set up the source amplitude accuracy test

To set up the source amplitude accuracy test

Performance Test and Operation Verification

This test verifies that the Agilent 35670A meets its source output specification for sine amplitude accuracy at 1 kHz. In this test, a digital multimeter measures the amplitude accuracy of the source. Source amplitude accuracy is checked at

0.1 Vpk with the digital multimeter set to the 100 mVrms range and at 3.0 and

5.0Vpk with the digital multimeter set to the 10 Vrms range. For the standard two channel analyzer, the digital multimeter is connected to the rear panel source connector instead of the front panel source connector. This is the only test that verifies the rear panel source port on the standard two channel analyzer.

1 2 ch

1 4 ch

3-45