To troubleshoot input and ADC failures

Troubleshooting the Analyzer

Agilent 35670A

To troubleshoot input and ADC failures

 

This is only a quick check of the Input assembly. If the Input assembly’s amplitude is still suspected of failing, set the analyzer to the failing range, impedance, and frequency. Connect a signal equal to the range setting to the failing channel. When the input level equals the range level, A1 P200 (channel 1) or A1 P700 (channel 2) should be 2.83 ±0.28 Vp-p.