Verifying Specifications

Agilent 35670A

Measurement Uncertainty

 

The sync output to signal output phase error was determined to be less than 0.25 degrees.

If measured value is within 3% of specification, verify synthesizer level accuracy. Note: Without 50 termination, observed levels are twice the setting into high impedance.

Performance Test

Using Recommended Test

Using Other Test Equipment

 

Equipment

 

 

 

 

 

 

 

 

 

 

 

Measurement

Ratio

Measurement

Ratio

 

Uncertainty

 

Uncertainty

 

 

 

 

 

 

 

 

 

 

 

Source Amplitude Accuracy

 

 

 

 

0.1 Vpk

± 9.83 mVpk

>10:1

 

 

3.0 Vpk

± 492.9 mVpk

>10:1

 

 

5.0 Vpk

 

 

± 633.0 mVpk

>10:1

 

 

 

 

 

 

 

 

 

 

Source Output Resistance

±0.15

>10:1

 

 

 

 

 

 

 

Source DC Offset

 

 

 

 

0 Vdc, 0 Vac-pk

± 238 nV

>10:1

 

 

±10 Vdc, 0 Vac-pk

± 84 mV

>10:1

 

 

0 Vdc, 5 Vac-pk

 

 

± 2.123 mV

>10:1

 

 

±5 Vdc, 5 Vac-pk

 

 

± 43 mV

>10:1

 

 

 

 

 

 

 

 

 

 

Source Flatness

± 0.2 dB

5.24:1

 

 

 

 

 

 

 

Source Distortion

 

 

 

 

fundamental <30 kHz

±0.5 dB

6.3:1

 

 

fundamental ≥30 kHz

±0.5 dB

>10:1

 

 

 

 

 

 

 

3-58

Page 116
Image 116
Agilent Technologies 35670-90066 manual Using Other Test Equipment