Agilent 35670A

 

 

Verifying Specifications

 

 

 

Performance Test Record - Two Channel

Serial Number:_______________________Report Number:____________________

 

Test Date:___/___/___

 

 

 

 

 

Measurement

Lower Limit

Upper Limit

Measured Value

Pass/Fail

 

 

(dB)

(dB)

(dB)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

11 dBVrms, 99.84 kHz, One Ch, Ch 1

0.2

0.2

 

 

 

 

 

 

 

 

 

27 dBVrms, 51.2 kHz, Two Ch, Ch 1

0.2

0.2

 

 

 

 

 

 

 

 

 

27 dBVrms, 51.2 kHz, Two Ch, Ch 2

0.2

0.2

 

 

 

 

 

 

 

 

 

9 dBVrms, 51.2 kHz, Two Ch, Ch 1

0.2

0.2

 

 

 

 

 

 

 

 

 

9 dBVrms, 51.2 kHz, Two Ch, Ch 2

0.2

0.2

 

 

 

 

 

 

 

 

 

11 dBVrms, 51.2 kHz, Two Ch, Ch 1

0.2

0.2

 

 

 

 

 

 

 

 

 

11 dBVrms, 51.2 kHz, Two Ch, Ch 2

0.2

0.2

 

 

Amplitude Linearity

Measurement

Lower Limit

Upper Limit

Measured Value

Pass/Fail

 

(dB)

(dB)

(dB)

 

 

 

 

 

 

 

 

 

 

 

13 dBVrms, Ch 1

0.0615

0.061

 

 

 

 

 

 

 

13 dBVrms, Ch 2

0.0615

0.061

 

 

 

 

 

 

 

1 dBVrms, Ch 1

0.105

0.104

 

 

 

 

 

 

 

1 dBVrms, Ch 2

0.105

0.104

 

 

 

 

 

 

 

15 dBVrms, Ch 1

0.33

0.318

 

 

 

 

 

 

 

15 dBVrms, Ch 2

0.33

0.318

 

 

 

 

 

 

 

29 dBVrms, Ch 1

1.551

1.316

 

 

 

 

 

 

 

29 dBVrms, Ch 2

1.551

1.316

 

 

 

 

 

 

 

43 dBVrms, Ch 1

13.823

5.088

 

 

 

 

 

 

 

43 dBVrms, Ch 2

13.823

5.088

 

 

 

 

 

 

 

53 dBVrms, Ch 1

30.116

10.896

 

 

 

 

 

 

 

53 dBVrms, Ch 2

30.116

10.896

 

 

7 of 14

Page 123
Image 123
Agilent Technologies 35670-90066 manual Amplitude Linearity, 11 dBVrms, 99.84 kHz, One Ch, Ch