Agilent 35670A

Verifying Specifications

 

 

Operation Verification Test Record - Four Channel

Serial Number:_______________________Report Number:____________________

Test Date:___/___/___

 

 

Two Ch, 44800 Hz Start, Ch 1

80

 

 

 

 

Two Ch, 44800 Hz Start, Ch 2

80

 

 

 

 

Two Ch, 46400 Hz Start, Ch 1

80

 

 

 

 

Two Ch, 46400 Hz Start, Ch 2

80

 

 

 

 

Two Ch, 48000 Hz Start, Ch 1

80

 

 

 

 

Two Ch, 48000 Hz Start, Ch 2

80

 

 

 

 

Two Ch, 49600 Hz Start, Ch 1

80

 

 

 

 

Two Ch, 49600 Hz Start, Ch 2

80

Amplitude Accuracy

Measurement

Lower Limit

Upper Limit

Measured Value

Pass/Fail

 

(dBVrms)

(dBVrms)

(dBVrms)

 

 

 

 

 

 

 

 

 

 

 

51 dBVrms, Ch 1

51.15

50.85

 

 

 

 

 

 

 

51 dBVrms, Ch 2

51.15

50.85

 

 

 

 

 

 

 

51 dBVrms, Ch 3

51.15

50.85

 

 

 

 

 

 

 

51 dBVrms, Ch 4

51.15

50.85

 

 

 

 

 

 

 

43 dBVrms, Ch 1

43.15

42.85

 

 

 

 

 

 

 

43 dBVrms, Ch 2

43.15

42.85

 

 

 

 

 

 

 

43 dBVrms, Ch 3

43.15

42.85

 

 

 

 

 

 

 

43 dBVrms, Ch 4

43.15

42.85

 

 

 

 

 

 

 

35 dBVrms, Ch 1

35.15

34.85

 

 

7 of 15

Page 167
Image 167
Agilent Technologies 35670-90066 manual Amplitude Accuracy