Agilent Technologies 35670-90066 manual To do the incoming inspection

Models: 35670-90066

1 439
Download 439 pages 40.89 Kb
Page 43
Image 43
To do the incoming inspection

Warning

Agilent 35670A

Preparing the Analyzer for Use

 

To do the incoming inspection

To do the incoming inspection

The Agilent 35670A Dynamic Signal Analyzer was carefully inspected both mechanically and electrically before shipment. It should be free of marks or scratches, and it should meet its published specifications upon receipt.

Inspect the analyzer for physical damage incurred in transit. If the analyzer was damaged in transit, do the following:

Save all packing materials.

File a claim with the carrier.

Call your Agilent Technologies sales and service office.

If the analyzer is mechanically damaged, the integrity of the protective earth ground may be interrupted. Do not connect the analyzer to power if it is damaged.

Check that the POWER SELECT switch on the analyzer’s rear panel is set to the AC position.

The switch is in the AC position when in the ‘’in’’ position.