Troubleshooting the Analyzer

Agilent 35670A

To troubleshoot source and calibrator failures

 

Step 4. Check the calibrator output.

GSet the power switch to off ( O ).

Remove the A1/A2 Input assembly and attach a test clip patch cord to TP 17. Connect a 10:1 oscilloscope probe to the patch cord and TP 8 (ground).

Reinstall the Input assembly in the card nest with patch cord and probe attacted.

Set the power switch to on ( l ).