To set up the source output resistance test

 

To set up the source output resistance test

Performance Test only

This test verifies that the Agilent 35670A meets its source output specification for resistance. In this test, a digital multimeter measures the 50 feedthrough termination. The channel 1 input then measures the source output across the feedthrough termination, then in an open circuit condition. The resistance is calculated using the following formula:

Rs = R1((Vopen - Vload)/Vload)

Note: Use the same 50 feedthrough termination for steps 1 and 2.

1 2 ch

2 2 ch

3-46