To adjust the ADC gain, offset and reference

Agilent 35670A

Adjusting the Analyzer

 

To adjust the ADC gain, offset and reference

To adjust the ADC gain, offset and reference

This procedure adjusts the second-pass gain, the first-pass offset, and the reference voltage for the ADC on the A5 Analog assembly. This prevents nonlinear Analog-to-Digital Converter (ADC) operation near the Digital-to-Analog Converter (DAC) transition levels.

Equipment Required: Oscilloscope

1:1 Oscilloscope Probe

Capacitive Load

BNC-to-BNC Cable

Set the power switch to off ( O ).

Connect the capacitive load (from the service kit) to the oscilloscope input. Connect the 1:1 oscilloscope probe to the capacitive load. Attach the probe to A5 TP400 and the probe ground clip to the instrument chassis.

Connect the oscilloscope’s external trigger connector to the analyzer’s SOURCE connector using a BNC cable.

5-7