Agilent Technologies 35670-90066 Binary, Hexa, ~Time LEDs, Description, DS5 DS9, decimal, Visible

Models: 35670-90066

1 439
Download 439 pages 40.89 Kb
Page 193
Image 193
Binary

Agilent 35670A

 

 

Troubleshooting the Analyzer

 

 

 

To troubleshoot power-up failures

Binary

Hexa-

~Time LEDs

Description

(DS5) (DS9)

decimal

Visible

 

 

 

 

 

 

 

 

 

1111 1111

FF

200 ms on

A7 flashes LEDs

0000 0000

00

200 ms off

 

 

 

 

 

0000 1000

08

starting A7 test

 

 

 

 

0000 0010

02

A8 RAM DSACK test

 

 

 

 

0001 0100

14

starting A8 RAM test

 

 

 

 

0001 0110

16

starting A8 refresh test

 

 

 

 

0001 1100

1C

4s

starting A8 program ROM test

 

 

 

 

0000 0000

00

4s

clear LEDs

 

 

 

 

1010 0000

A0

A7 MFP test

 

 

 

 

1010 0001

A1

starting A7 DSP test

 

 

 

 

1010 0010

A2

fast bus test

 

 

 

 

0101 1110

AE

200 ms

front panel test

 

 

 

 

1111 1111

FF

Remain on

 

0 = LED off

1 = LED on

When no failure occurs, these codes appear for only a very short time and probably won’t be visible.

If the power-on LEDs display a code for more than 4 seconds, a failure occurred in the core assemblies or on the buses.

For additional information on the power-on test, see the ‘’Power-on Test Descriptions’’ on page 10-3.

Step 3. Determine the next step by comparing the power-on test results to the following table.

G

Power-on LEDs

Next Test

 

 

 

 

LEDs stop and display a fail code.

CPU, memory, and

A7 DS101 (green, run LED) is off.

buses, page 4-18

 

 

LEDs pass, but the screen is defective.

Display, 4-22

 

LEDs pass, but it takes more than 3 minutes before the keys are active. IIC Bus, page 4-25

LEDs pass and screen appears normal, but keys do not function.

 

4-17

Page 193
Image 193