Adjusting the Analyzer

Agilent 35670A

To adjust filter flatness

 

Press the following keys:

[ Rtn ]

[ CHANNEL 3 ADJUSTMNT ] [ 25 kHz FLATNESS ]

Wait for the analyzer to set up the adjustment. The analyzer is ready when the adjustment message appears on the screen.

While monitoring the Yr: value, adjust A2 R615 in the lower slot for a marker value of 0 0.1 dB.

Press the following keys:

[ Rtn ]

[ CHANNEL 4 ADJUSTMNT ] [ 25 kHz FLATNESS ]

Wait for the analyzer to set up the adjustment. The analyzer is ready when the adjustment message appears on the screen.

While monitoring the Yr: value, adjust A2 R615 in the upper slot for a marker value of 0 0.1 dB.