To troubleshoot trigger failures

Troubleshooting the Analyzer

Agilent 35670A

To troubleshoot trigger failures

 

To troubleshoot trigger failures

Use this test when the trigger mode is suspected of failing or the Input Trigger self test fails on all channels.

Step 1. Check trigger modes.

GSet the power switch to on ( l ).

Press the following keys:][System UtilityCALIBRATN ]AUTO CAL OFF[ Input ][ [[[[]]]1 ALL CHANNELS ]CH* FIXED RANGEVpk

[ Source ]

[ SOURCE ON ] [ LEVEL ]

1

[ Vpk ]

[ FIXED SINE ] [ 1 ]

[ kHz ]

[ Inst Mode ]

[ 4 CHANNEL ] or [ 2 CHANNEL ]

[ ACTIVE TRACE ] [ A B C D ] or [ A B ]

[ Meas Data ]

[ ALL CHANNELS ]

[ TIME CHANNEL ]

[ System Utility ]

[ MORE ]

[ SERVICE TESTS ]

[ SPCL TEST MODES ] [ SOURCE LEVEL ]

[ Trigger ]

[ SOURCE TRIGGER ]

If the analyzer is not triggering and the message WAITING FOR SOURCE TRIGGER is displayed, the A6 Digital assembly is probably faulty.

Fixed sine wave source triggering occurs at a consistent (but not predictable) point within the time record.

Set the frequency synthesizer as follows:

Frequency

1 kHz

Amplitude

6 Vp-p

Function

Square Wave

4-62