Verifying Specifications

Agilent 35670A

Measurement Uncertainty

 

Measurement Uncertainty

The following table lists the measurement uncertainty and ratio for each performance test using the recommended test equipment. Except for the External Trigger test, the ratios listed for the recommended test equipment meet or exceed the measurement uncertainty ratio required by U.S. MIL-STD-45662A.

If you are using equipment other than the recommended test equipment, you may calculate and record the measurement uncertainty and ratio for each performance test. The table may be reproduced without written permission of Agilent Technologies.

Performance Test

Using Recommended Test

Using Other Test Equipment

 

 

 

Equipment

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measurement Uncertainty

Ratio

Measurement Uncertainty

Ratio

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Self Test

 

 

NA

NA

NA

NA

 

 

 

 

 

 

 

DC Offset

 

 

NA

NA

NA

NA

 

 

 

 

 

 

 

Noise

 

 

NA

NA

NA

NA

 

 

 

 

 

Spurious Signals

NA

NA

NA

NA

 

 

 

 

 

Amplitude Accuracy

 

 

 

 

51 dBVrms

±0.020 dB

7.7:1

 

 

43 dBVrms

±0.0084 dB

>10:1

 

 

35 dBVrms

±0.004 dB

>10:1

 

 

>10:1

 

 

27 dBVrms

±0.003 dB

 

 

>10:1

 

 

11 dBVrms

±0.001 dB

 

 

>10:1

 

 

1 dBVrms

 

 

±0.0008 dB

 

 

 

 

>10:1

 

 

9 dBVrms

 

 

±0.001 dB

 

 

 

 

>10:1

 

 

19 dBVrms

 

±0.00081 dB

 

 

 

>10:1

 

 

27 dBVrms

 

 

 

 

±0.00117 dB

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Flatness

 

 

 

 

 

 

25.6 kHz, 27 dBVrms

±0.01487 dB

>10:1

 

 

25.6 kHz, 9 dBVrms

±0.01277 dB

>10:1

 

 

25.6 kHz,

11 dBVrms

±0.01277 dB

>10:1

 

 

 

10:1

 

 

51.2 kHz, 27 dBVrms

±0.02025 dB

 

 

51.2 kHz, 9 dBVrms

>10:1

 

 

±0.01460 dB

 

 

51.2 kHz, 11 dBVrms

>10:1

 

 

±0.01583 dB

 

 

99.84 kHz, 27 dBVrms

10:1

 

 

±0.02025 dB

 

 

 

 

 

99.84 kHz, 9 dBVrms

>10:1

 

 

±0.01460 dB

 

 

 

 

 

99.84 kHz, 11 dBVrms

>10:1

 

 

±0.01583 dB

 

 

 

 

 

 

 

 

NA (not applicable) internal test

3-56

Page 114
Image 114
Agilent Technologies 35670-90066 manual Measurement Uncertainty