Agilent Technologies 35670-90066 manual Symptom, Probable Faulty Assembly

Models: 35670-90066

1 439
Download 439 pages 40.89 Kb
Page 182
Image 182
Probable Faulty Assembly

Troubleshooting the Analyzer

Agilent 35670A

To perform initial verification

 

Step 4. If the analyzer powers up normally with no error messages (see the following

Gillustration), the screen is continually updating, but the analyzer does not respond to key presses, use the following table to determine the probable faulty assembly.

Symptom

Probable Faulty Assembly

 

 

 

 

SYSTEM keys function but MARKER, DISPLAY,

A13 or A15 Primary Keypad

MEASUREMENT, or number keys do not function correctly

 

 

 

MARKER, DISPLAY, MEASUREMENT, and number keys

A14 Secondary Keypad

function but SYSTEM keys do not function correctly

 

 

 

No keys function correctly

A11 Keyboard Controller

4-6

Page 182
Image 182
Agilent Technologies 35670-90066 manual Symptom, Probable Faulty Assembly