Agilent 35670A

Troubleshooting the Analyzer

 

To perform self tests

To perform self tests

Use this test when one of the following occurs:

 

 

Performance test fails

 

 

Calibration fails

 

 

Trigger fails

 

 

GPIB fails

 

 

Microphone power fails

 

 

Serial port fails

 

 

Parallel port fails

 

 

Failure is intermittent

 

 

Step 1. Run all the functional self tests.

 

 

 

Connect the rear panel SOURCE output to the rear panel TACH input

 

 

 

using a BNC cable.

 

 

 

Remove all cables from the front panel input connectors.

 

 

 

Caution

G

The ICP self test outputs approximately 30 Vdc on the input connectors. Before

 

starting the self tests, disconnect all devices connected to the input connectors.

 

Devices left connected during the ICP self test may be damaged.

 

 

Set the power switch to on ( l ).

 

 

 

After calibration is complete, press the following keys:

[ Input ]

[ ALL CHANNELS ]

[ CH* FIXED RANGE ] [ System Utility ]

[ CALIBRATN ]

[ AUTO CAL OFF ] [ Rtn ]

[ MORE ]

[ SELF TEST ]

[ FUNCTIONL TESTS ] [ ALL ]

[ CONTINUE ]

4-31

Page 207
Image 207
Agilent Technologies 35670-90066 manual To perform self tests, Using a BNC cable, Input