Circuit Descriptions

Agilent 35670A

Option UK4 Microphone Adapter and Power Supply

 

 

Option UK4 Microphone Adapter and Power Supply

 

The optional Microphone Adapter and Power Supply provides four LEMO connectors

 

with power for microphones. The input signal from each LEMO connector is routed

 

to a BNC connector. BNC cables then connect the input signals to the analyzer’s input

 

channels.

+5 V Regulator

Regulates +8 V to +5 V.

DC-to-DC Converter

Converts +5 V to +28 V. The second DC-to-DC Converter converts +28 V to +200 V.

Filter

Filters +28 V providing the microphone preamplifier voltage to pins 5 and 6 of each LEMO

 

connector.

On Off

Connects a 200 V polarization voltage or ground to pin 3 of each LEMO connector.

Heater Input Power

Allows an externally supplied heater voltage to be connected to pin 1 of each LEMO

 

connector.

LED Drivers

Provides control lines for the overload and measurement LEDs on the HP 35230A Sound

 

Intensity Probe.

Buffer

Buffers the trigger signal from the HP 35230A Sound Intensity Probe. A BNC cable connects

 

the trigger signal to the analyzer’s rear panel.

Sound Intensity Probe

Provides the interface to the HP 35230A Sound Intensity Probe.

Connector

 

8-40

Page 342
Image 342
Agilent Technologies 35670-90066 manual Option UK4 Microphone Adapter and Power Supply