Agilent 35670A

 

Troubleshooting the Analyzer

 

 

To troubleshoot CPU, memory, and buses failures

Binary

Hexadecimal

Probable Faulty Assembly

(DS5) (DS9)

 

 

 

 

 

 

 

 

0000 0100

04

A7 CPU

1111 1111

FF

 

0001 0011

13

 

0000 0001

01

 

0001 0111

17

 

0001 1000

18

 

0000 1001

09

 

0000 1011

0B

 

0001 1010

1A

 

0001 1001

19

 

0000 1000

08

 

0001 0010

12

 

1010 0000

A0

 

 

 

 

0000 0010

02

A8 Memory

0001 1011

1B

 

0001 0100

14

 

0001 0110

16

 

0001 1100

1C

 

0 = LED off

1 = LED on

Step 2. Determine if the CPU assembly is causing the failure.

GSet the power switch to off ( O ).

Pull the following assemblies out of the card nest about 1 inch:A6 DigitalA5 AnalogA2 Input (optional)A1 or A2 Input

4-19

Page 195
Image 195