A10 Rear Panel

 

Agilent 35670A

Voltages and Signals

 

 

A10 Rear Panel

DIO1 — DIO8

Data Input/Output — These are inverted data lines that conform to IEEE specification

 

IEEE-488. When ATN is low, these lines contain interface commands. When ATN is high,

 

these lines contain data.

 

EOIn

End or Identify — If ATN is high, a low on this line marks the end of a message block. If

 

ATN is low, a low on this line requests a parallel poll.

 

IFCn

Interface Clear — This line goes low to halt all current operations on the bus, unaddress all

 

other devices, and disable serial poll. The system controller becomes the controller in charge.

 

This line is only used during the analyzer’s developement.

 

NDACn

Not Data Accepted — This line goes high when the DIO lines have been latched by

 

the acceptor.

 

NRFDn

Not Ready for Data — This line goes high when the acceptor is ready to accept data.

RENn

Remote Enable — This line is low when the GPIB has control and high during local operation.

SRQn

Service Request — This line is low when a device on the GPIB needs service.

 

Serial Port

The following table lists signals at the Serial Port connector (A10 J101). The Serial Port is a 9-pin EIA-574 port that is only available using Instrument Basic. A description of each signal follows the table.

Signal Name

Pin

 

 

DSR

6

 

 

DTR

1

 

 

RTS

4

 

 

RxD

3

 

 

TxD

2

 

 

Logic Gnd

7

 

 

Not Used

5, 8-9

DSR

Data Set Ready — Some devices check this line for a high to verify that the analyzer is

 

connected and ready. The user can set this line high or set this line to go high only when the

 

analyzer is ready for data transfer.

DTR

Data Terminal Ready — This line is tied high. Some devices check this line for a high to

 

verify that the analyzer is connected and ready.

RTS

Request To Send — This line is tied high. Some devices require this line to be high before

 

transferring data.

RxD

Receive Data — This is the serial EIA-574 receive data line. This line transmits data from

 

peripheral devices one byte at a time.

TxD

Transmit Data — This is the serial EIA-574 transmit data line. This line transmits data to

 

peripheral devices one byte at a time.

9-15