Agilent 35670A

Voltages and Signals

 

 

A100 Disk Drive

DIR

Direction — This line sets the direction for the disk head. A high on this line sets the direction

 

away from the spindle. A low on this line sets the direction toward the spindle.

DISKINn

Disk In — This line goes low when a flexible disk is inserted in the A100 Disk Drive assembly.

DRIVESELn

Drive Select — A low on this line selects the A100 Disk Drive assembly.

 

DSKCHGn

Disk Change — This line goes low when a flexible disk is removed from the A100 Disk Drive

 

assembly. This line remains low until a flexible disk is installed and STEPn goes low.

HDSEL

Head Select — A low on this line selects the lower disk drive head. A high on this line selects

 

the upper disk drive head.

 

HIDENSn

High Density Select — This line goes low when a high density flexible disk is inserted in the

 

A100 Disk Drive assembly.

 

INDEXn

Index — This line pulses low with each revolution of the flexible disk.

 

MTRn

Motor On — A low on this line turns on the disk drive motor.

 

ReDATAn

Read Data — This line pulses low for each bit detected on the flexible disk.

 

SEL0 - SEL1

Drive Select — When SEL0 and SEL1 are low, the A100 Disk Drive assembly is enabled.

STEPn

Step — A low on this line moves the disk drive head. When STEPn and DIR are low, the head

 

moves toward the disk spindle. When STEPn is low and DIR is high, the head moves away

 

from the disk spindle.

 

T00n

Track 00 — This line is low when the head is positioned over track 0 on the flexible disk.

WDATA

Write Data — When WGATE is low, a low pulse on this line writes a bit to

 

 

the disk.

 

WGATE

Write Gate — When this line is low, information may be written to the A100 Disk Drive

 

assembly under control of the WDATA line.

 

WRIPROTn

Write Protect — This line is low when a write-protected disk is installed in the A100 Disk

 

Drive assembly.

 