Agilent Technologies 35670-90066 Recommended Test Equipment, Instrument, Critical Specifications

Models: 35670-90066

1 439
Download 439 pages 40.89 Kb
Page 35
Image 35
Recommended Test Equipment

Agilent 35670A

Specifications

 

Recommended Test Equipment

Recommended Test Equipment

The following table lists the recommended equipment needed to test the performance of the Agilent 35670A Dynamic Signal Analyzer. The table on page 1-20 lists additional equipment needed to adjust and troubleshoot the analyzer. Other equipment may be substituted for the recommended model if it meets or exceeds the listed critical specifications. When substitutions are made, you may have to modify the procedures to accommodate the different operating characteristics.

Recommended Test Equipment

Instrument

Critical Specifications

Recommended Model

 

 

 

AC Calibrator 10 Hz to 102.4 kHz; 1 mV to 10 V Amplitude

Amplitude Accuracy: ±0.1% phase locking capability

Fluke 5700A † Alternate

Fluke 5200A †

Datron 4200, 4700, or 4708 ‡ HP 745A

Frequency Synthesizer

Frequency Range: 10 Hz to 1 MHz

Frequency Accuracy: 5 ppm Amplitude Accuracy:

0.2dB from 1 Hz to 100 kHz 1 dB from 100 kHz to 1 MHz

Harmonic Distortion: –70 dBc

Spurious: –70 dBc

<±1 deg phase shift between output and sync

HP 3326A

Alternate

(2) HP 3325A/B Opt 001

Low Distortion

Frequency Range: 10 Hz to 100 kHz

HP 339A ††

Oscillator

Harmonic Distortion: –93 dB, 10 Hz to

Alternate

 

20 kHz

HP 3326A with notch filter ††

 

 

HP 3325A/B with notch

 

 

filter††

 

 

 

Digital Multimeter

5 1/2 digit True rms ac Voltage:

30 Hz to 100 kHz; 0.1 to 500 V; ±0.1%; 1 Minput impedance dc Voltage:

1 V to 300 V; ±0.1%

HP 3458A

Alternate

HP 3456A , HP 3455A HP 3478A

Feedthrough

50 : ±2% at dc

Pomona Elect Model 4119-50

Termination (2)

 

‡‡

(4 for option

 

Alternate

AY6)

 

HP 11048C, HP 10100C

John Fluke Manufacturing Co., Inc., PO Box C9090, Everett, WA 98206 U.S.A. (206) 347-6100

‡ Wavetek, 5808 Churchman Bypass, Indianapolis, IN 46203 U.S.A.

†† This equipment is not required for Operation Verification. The parts and schematic for the notch filter are shown on page 1-19.

‡‡ ITT Pomona Electronics, 1500 East Ninth Street, Pomona, CA 91769 U.S.A. (714) 469-2900 FAX (206) 629-3317

1-17