Agilent 35670A

Troubleshooting the Analyzer

 

To troubleshoot source and calibrator failures

Press the following keys:

[ System Utility ]

[CALIBRATN ]

[ AUTO CAL OFF ]

[Input ]

[ ALL CHANNELS ]

[ CH* FIXED RANGE ] 1

[ Vpk ]

[ Source ]

[SOURCE ON ]

[LEVEL ] 1

[ Vpk ]

[ System Utility ]

[MORE ]

[ SERVICE TESTS ]

[ SPCL TEST MODES ] [ HIGH LEVEL CAL ]

If the signal does not look like the following figure, the A5 Analog assembly is probably faulty.

Oscilloscope Setup

 

Parameters

Waveform

 

 

 

 

 

 

 

 

Connect CH1 to A1/A2 TP 17

Amplitude

 

 

 

 

CH1 V/div

2 V/div

Time

 

 

 

 

 

Input

1 M Ω

 

 

 

Impedance

dc

 

 

 

CH1 Coupling

10

 

 

 

Probe Atten

Repetitive

 

 

 

Display Mode

8

 

 

 

Averaging

20 ms/div

 

 

 

Time/div

Trg’d Sweep

 

 

 

Trigger

Chan1

 

High Level Calibrator

Trig Src

 

 

 

 

 

 

Press [ LOW LEVEL CAL ].

If the signal does not look like the following figure, the A5 Analog assembly is probably faulty.

Oscilloscope Setup

Parameters

Waveform

 

 

 

 

 

 

 

Connect CH1 to A1/A2 TP 17 Amplitude

 

 

 

 

CH1 V/div

Time

 

 

200 mV/div

 

 

Input

1 M Ω

 

 

Impedance

dc

 

 

CH1 Coupling

10

 

 

Probe Atten

Repetitive

 

 

Display Mode

8

 

 

Averaging

20 ms/div

 

 

Time/div

Trg’d Sweep

 

 

Trigger

Chan1

 

 

Low Level Calibrator

 

Trig Src

 

 

 

 

 

4-49