Troubleshooting the Analyzer

Agilent 35670A

To troubleshoot self-test lockup failures

 

Step 2. Run the IIC and fast bus self tests.

GPress the following keys:

[ System Utility ]

[ CALIBRATN ]

[ AUTO CAL OFF ] [ Input ]

[ ALL CHANNELS ]

[ CH* FIXED RANGE ] 1

[ Vpk ]

[ System Utility ] [ MORE ]

[ SELF TEST ] [ TEST LOG ]

[ Rtn ]

[ FUNCTIONL TESTS ] [ I/O ]

[ IIC BUS ]

If the keyboard is not active, or the analyzer locks up when a key is pressed, go to page 4-25,‘’To troubleshoot IIC bus failures.’’

Press the [ FAST BUS ] softkey.

If the analyzer locks up or the digital processor failed the fast bus self test, go to page 4-29,‘’To troubleshoot fast bus failures.’’

4-38