Verifying Specifications

Agilent 35670A

To run the program without a printer

 

To run the program without a printer

Use this procedure if you do not have an GPIB printer connected to yout system.

Write in the information needed on the title page of the selected test record.

The test records are located near the back of this chapter and may be copied without written permission of Agilent Technologies.

Press the following keys and when the program prompts you, type in the model number and GPIB address:

[ EQUIP CONFIG ] [ AC CALIBRATO ] [ SYNTH. 1 ]

[ SYNTH. 2 ] (If needed)

[ LOW-D OSCILLATO ] (If needed)

[ MULTIMETER ] [ RETURN ]

The GPIB addresses equals 100 (interface select code) + (primary address). The interface select code for the test equipment is 7 (for example, if the primary address is 8, the GPIB address is 708).

Press the following keys:

[ TEST CONFIG ]

[ PROCEDURE ]

[ OP_VERIFY ] or [ PERFORMAN ]

[ STOP AFTER ]

[ EACH MEASUREMENT ] [ RETURN ]

Press the following keys to start the test:

[ START TESTING ]

[ START BEGINNING ]

3-10