Agilent Technologies 35670-90066 manual To remove CPU

Models: 35670-90066

1 439
Download 439 pages 40.89 Kb
Page 281
Image 281
To remove CPU

Agilent 35670A

Replacing Assemblies

 

To remove CPU

To remove CPU

1 Remove cover (see ‘’To remove cover’’).

2 Using a T-10 torx driver, remove the nine screws from the A7 CPU assembly. Lift the assembly up, unpluging the A7 CPU assembly from the A8 Memory assembly and A99 Motherboard.

3 Disconnect the ribbon cables from the A7 CPU assembly.

4 The analyzer’s serial number and firmware options are stored in EEPROM (U27) on the A7 CPU assembly. Before replacing the CPU assembly, remove A7 U27 from the faulty assembly and insert into the new assembly. See ‘’What to do before replacing the CPU assembly’’ on page 6-3.

6-11

Page 281
Image 281
Agilent Technologies 35670-90066 manual To remove CPU