To adjust the source

Adjusting the Analyzer

Agilent 35670A

To adjust the source

 

To adjust the source

This procedure adjusts the source dc offset on the A5 Analog assembly.

Equipment Required: Multimeter

BNC-to-Dual Banana Cable

Connect the multimeter to the analyzer’s rear-panel SOURCE connector.

Press the following keys:

[ System Utility ]

[ MORE ]

[ SERVICE TESTS ] [ ADJUSTMTS ]

[ SOURCE ADJUSTMNT ] [ DC OFFSET ]

Adjust A5 R48 for 0 Vdc±1 mV.

Press the [ FILTER DC OFFSET ] softkey.

Adjust A5 R59 for 0 Vdc ±1 mV.