Agilent 35670A

Troubleshooting the Analyzer

 

To troubleshoot trigger failures

If the signal at A5 TP 204 is toggling, the A6 Digital assembly is probably faulty.

If the signal at A5 TP 204 is not toggling, the A5 Analog assembly is probably faulty.

4-65

Page 241
Image 241
Agilent Technologies 35670-90066 manual Agilent 35670A