Adjusting the Analyzer

This chapter contains the adjustment procedures for the Agilent 35670A Dynamic Signal Analyzer. Use these adjustments if the analyzer does not meet its specifications or if instructed in chapter 4, ‘’Troubleshooting the Analyzer,’’ or chapter 6, ‘’Replacing Assemblies,’’ to perform these adjustments. These adjustments are not required for routine maintenance.

Allow the Agilent 35670A analyzer to warm up for an hour before doing any of the adjustments.

During many of these adjustment procedures, an adjustment message appears on the screen. The instructions on the screen are not as complete as the instructions in this guide. When an adjustment message appears on the screen, continue to follow the instructions in this guide. Failure to follow the instructions in this guide may result in an incorrect adjustment, which would appear as a hardware failure.

The following table shows the assembly and components adjusted during each adjustment procedure.

Adjustment

Assembly

Component

 

 

 

Frequency reference

A7 CPU

A7 C85

 

 

 

Source

A5 Analog

A5 R48

 

 

A5 R59

 

 

 

ADC gain, offset and

A5 Analog

A5 R407

reference

 

A5 R405

 

 

A5 R431

 

 

 

Input dc offset

A1/A2 Input

A1/A2 R39

 

 

A1/A2 R539

 

 

 

Common mode rejection

A1/A2 Input

A1/A2 R43

 

 

A1/A2 R543

 

 

 

Filter flatness

A1/A2 Input

A1/A2 R115

 

 

A1/A2 R235

 

 

A1/A2 R615

 

 

 

Display voltage

A102 DC-DC Converter

A102 R25

5-2

Page 250
Image 250
Agilent Technologies 35670-90066 manual Adjusting the Analyzer, Adjustment Assembly Component