Troubleshooting the Analyzer

Agilent 35670A

To troubleshoot auto-range failures

 

Press the following keys:

[ CHANNEL 1 ]

[ CHANNEL 1 RANGE ]

The range should be set to 5 dBVrms.

Press [ Rtn ].

Repeat steps 5 and 6 for each channel.

If only one channel, or channel 1 and 3, or channel 2 and 4 are failing, the A1/A2 Input assembly is probably faulty.

In the two channel analyzer, the A1 Input assembly provides the circuits for channel 1 and 2. In the four channel analyzer, the A2 Input assembly in the lower slot provides the circuits for channel 1 and 3, and the A2 Input assembly in the upper slot provides the circuits for channel 2 and 4.

If all channels are failing, the A5 Analog assembly is probably faulty.

4-60

Page 236
Image 236
Agilent Technologies 35670-90066 manual Range should be set to 5 dBVrms