Troubleshooting the Analyzer

 

Agilent 35670A

To perform self tests

 

 

Self-Test Troubleshooting Guide (continued)

 

Failing Self Test

Probable Faulty Assembly

Adjustment

Troubleshooting Test

 

 

 

 

 

 

 

 

 

 

Source With LO fails and

A6

Digital

 

 

Source Without LO passes

 

 

 

 

 

 

 

 

Source Without LO

A1/A2 Input

 

Input and ADC, page 4-51

one channel

A5

Analog

 

Four channel, page 4-54

 

 

 

 

 

Source Without LO

A2

Input

 

 

channel 1 and 3 or

 

 

 

 

channel 2 and 4

 

 

 

 

 

 

 

 

Source With LO

A1/A2 Input

 

Input and ADC, page 4-51

one channel

A5

Analog

 

Four channel, page 4-54

 

 

 

 

 

Source With LO

A2

Input

 

 

channel 1 and 3 or

 

 

 

 

channel 2 and 4

 

 

 

 

 

 

 

 

Input Offset

A1/A2 Input

Input dc offset, page 5-10

 

one channel or

 

 

 

 

channel 1 and 3 or

 

 

 

 

channel 2 and 4

 

 

 

 

 

 

 

 

 

ADC Gate Array

A5

Analog

ADC gain, offset, and

Source and calibrator,

 

A6

Digital

reference, page 5-10

page 4-45

 

 

 

 

 

Source Without LO

A5

Analog

 

Source and calibrator,

all channels

A6

Digital

 

page 4-45

 

 

 

 

 

Source With LO

A5

Analog

 

Source and calibrator,

all channels

A6

Digital

 

page 4-45

 

 

 

 

 

Input Offset

A5

Analog

ADC gain, offset, and

 

all channels

 

 

reference, page 5-7

 

 

 

 

 

Input Distortion

A1/A2 Input

 

Input and ADC, page 4-51

one channel

A5

Analog

 

Four channel, page 4-54

 

 

 

 

 

Input Distortion

A2

Input

 

 

channel 1 and 3 or

 

 

 

 

channel 2 and 4

 

 

 

 

 

 

 

 

 

Input Distortion

A5

Analog

ADC Gain, offset, and

Source and calibrator,

all channels

A6

Digital

reference, page 5-7

page 4-45

 

 

 

 

 

Input Trigger

A5

Analog

 

 

one or more channels fail,

 

 

 

 

but at least one channel

 

 

 

 

passes

 

 

 

 

 

 

 

 

 

Input Trigger

A5

Analog

 

Trigger, page 4-62

 

A6

Digital

 

 

4-34

Page 210
Image 210
Agilent Technologies 35670-90066 manual Troubleshooting the Analyzer