•Remove the A7 CPU assembly.

Agilent 35670A

Troubleshooting the Analyzer

 

To troubleshoot the power supply

Remove the A7 CPU assembly.

See ‘’To remove CPU’’ on page 6-11.

Set the power switch to on ( l ).

If the power supply LED is still off, set the power switch to off ( O ), reconnect the CPU assembly, and go to Step 6.

Step 5. Repeat the following steps until the assembly causing the Power Supply

Gassembly to shut down is located.

Set the power switch to off ( O ).

Reconnect one assembly at a time in the following order:

A7 CPU (A7 P10 to A99 J7)

A8 Memory (A8 P1 to A7 J3)

A102 DC-DC Converter (cable to A7 P2)

A11 Keyboard Controller (cable to A7 P1)

A100 Disk Drive (cable to A7 P3)

Set the power switch to on ( l ).

If the green power supply LED is off, the assembly just reconnected is probably faulty.

A7 Component Locator, Circuit Side

4-13